Table 1.
The dose comparison between ionization chamber and TLDs.
| D TLD (cGy) | D IC (cGy) | Difference (%) | |
|---|---|---|---|
| 6 MV photon | 194.48 ± 2.32 | 199.50 ± 1.92 | −2.519% |
| 10 MV photon | 197.21 ± 3.93 | 197.97 ± 1.87 | −0.386% |
| 6 MeV electron | 203.71 ± 2.04 | 199.32 ± 1.56 | 2.205% |
| 9 MeV electron | 194.84 ± 3.62 | 198.45 ± 2.03 | −1.819% |
| 12 MeV electron | 198.99 ± 3.13 | 198.04 ± 1.58 | 0.478% |
D TLD: dose measured from TLDs; D IC: dose measured from the ionization chamber.