Published in final edited form as: Phys Med Biol. 2013 Oct 8;58(21):7463–7479. doi: 10.1088/0031-9155/58/21/7463

Table 1.

Characteristics of each field for the 30.6-Gy, 6-MV CSI treatment. For easier readability, the characteristics of the fields of feathered junctions and field-in-field fields were averaged (Y-jaw settings) or summed (monitor units) in this table.

AUBMC Fields
Description Right Cranial Left Cranial Upper Spinal Lower Spinal
Monitor units 1701 1665 3374 3656
Gantry angle (°) 90 270 0 0
Collimator angle (°) 5.0 355.0 0 0
Dose to calculation point (Gy) 15.3 15.3 30.6 30.6
X1 jaw (cm) −11.4 −9.0 −3.6 −4.0
X2 jaw (cm) 9.0 11.4 3.1 5.0
Y1 jaw (cm) −9.3 −9.3 −12.4 −17.7
Y2 jaw (cm) 15.1 15.1 9.6 15.0
Water equivalent depth of calculation point (cm) 7.64 7.86 3.74 5.66
Effective square (cm) 15.3 15.3 9.0 10.5
Source-to-skin distance (cm) 93.4 92.0 100.0 100.0
MD Anderson Fields
Description Right Cranial Left Cranial Upper Spinal Lower Spinal
Monitor units 1700 1700 3009 3213
Gantry angle (°) 270 90 0 0
Collimator angle (°) 8.0 352.0 0 0
Dose to calculation point (Gy) 15.4 15.4 32.6* 32.4*
X1 jaw (cm) −12.0 −12 −4.2 −5.3
X2 jaw (cm) 12.0 12 4.3 4.8
Y1 jaw (cm) −13.0 −13.0 −13.5 −14.0
Y2 jaw (cm) 11.0 11.0 14.6 14.6
Water equivalent depth of calculation point (cm) 10.3 10.3 4.2 6.7
Effective square (cm) 15.0 15.0 13.0 14.9
Source-to-skin distance (cm) 92.1 92.1 100 100
*

The spinal fields were normalized to deliver a DRx of 30.6 Gy before the field-in-field fields were added to the plan. The addition of these fields resulted in a 6% increase in the absorbed dose at the calculation points of the spinal fields for the MD Anderson plan. This did not affect the model results because the model was reported in absorbed dose per prescribed dose.