Figure 1. 3D sequence for combined phase-sensitive LGE imaging and T1 mapping.

Figure 1

Four sets of 3D imaging volumes are acquired using hybrid saturation and inversion pre-pulses with different inversion times (Tinvnull,Tinv1,Tinv2, and Tinv3). Imaging data acquired with Tinvnull will be used for reconstruction of the 3D LGE image therefore the Tinvnull time is selected to null the healthy myocardium. Tinv13 times are spread across the RR-interval and the data acquired with these inversion times in addition to the Tinvnull will be used for calculation of 3D T1 maps. To obtain fully-sampled LGE data, we propose to acquire data with Tinvnull in every other heart-beat and acquiring the Tinv13 data in between. This will result in fully sampled LGE data and 3× undersampling of the supplementary images used for T1 mapping.