Table 1.

T1 accuracy measured in a phantom experiment using the proposed LGE/T1 sequence, defined as the absolute difference of the mean T1 time in each vial and the spin-echo reference, averaged over ten repetitions. The proposed sequence was performed with different inversion-times in the LGE image, resulting in nulling of the different vials. Each sequence was repeated 10 times. The reference T1 times, assessed with an inversion recovery spin-echo sequence are given in parentheses.

Accuracy
LGE/T1 Vial #1
(271 ms)
Vial #2
(327 ms)
Vial #3
(420 ms)
Vial #4
(583 ms)
nulling: Vial #1 2.7 5.6 2.8 9.4
nulling: Vial #2 2.6 5.9 3.1 9.2
nulling: Vial #3 1.6 5.3 4.0 9.1
nulling: Vial #4 2.9 6.3 4.2 10.4