Skip to main content
. 2015 Apr 17;10(4):e0122433. doi: 10.1371/journal.pone.0122433

Table 2. Mean measurements, standard deviation and coefficient of variation of 11 daily testing of quality control reagent per reagent and per Nova StatSensor Xpress Creatinine analyzer.

  StatSensor 1 StatSensor 2 StatSensor 3
  mean SD CV % mean SD CV % mean SD CV %
Level 1 97.4 5.64 5.8 96.7 4.84 5.0 92.2 5.93 6.4
Level 2 210.3 12.9 6.1 211.5 9.89 4.7 200.9 18.1 9.0
Level 3 597.6 26.7 4.5 611.6 25.7 4.2 583.0 37.7 6.5