Skip to main content
. 2015 Apr 17;10(4):e0122433. doi: 10.1371/journal.pone.0122433

Table 3. Results of the measurements of the 5 linearity reagents per Nova StatSensor Xpress Creatinine analyzer and corresponding correlation coefficient.

StatSensor 1 StatSensor 2 StatSensor 3
Creatinine concentration (μmol/L)
88 88 99 99
203 229 214 179
398 378 415 399
601 588 606 623
866 895 908 804
Concordance coeffficient 0.997 0.997 0.993