Table 7. Comparison of creatinine clearance results determined by Nova StatSensor Xpress Creatinine analyzers or reference method modeled from the results obtained in the study applied to a hypothetical 50-kg, 20 year old female (N = 60).

    Nova StatSensor Xpress Creatinine  
    <10 mL/min 10–30 mL/min 30–50 mL/min ≥50 mL/min Total
Reference method <10 mL/min 3 2 0 0 5
10–30 mL/min 0 25 0 0 25
30–50 mL/min 0 2 6 0 8
≥50 mL/min 0 0 2 20 22
  Total 3 29 8 20 60