Table 7. Comparison of creatinine clearance results determined by Nova StatSensor Xpress Creatinine analyzers or reference method modeled from the results obtained in the study applied to a hypothetical 50-kg, 20 year old female (N = 60).
Nova StatSensor Xpress Creatinine | ||||||
---|---|---|---|---|---|---|
<10 mL/min | 10–30 mL/min | 30–50 mL/min | ≥50 mL/min | Total | ||
Reference method | <10 mL/min | 3 | 2 | 0 | 0 | 5 |
10–30 mL/min | 0 | 25 | 0 | 0 | 25 | |
30–50 mL/min | 0 | 2 | 6 | 0 | 8 | |
≥50 mL/min | 0 | 0 | 2 | 20 | 22 | |
Total | 3 | 29 | 8 | 20 | 60 |