Table 2.

Examples of SWI and QSM Protocols for the Brain

Field Recon Resolution (mm3) Flip angle TE1 (msec) TR (msec) # echo
1.5 T SWI 0.8 × 0.8 × 2 20° 40 50 1
QSM 0.8 × 0.8 × 2 30° Min 58 Multi
3.0 T SWI 0.8 × 0.8 × 2 15° 20 28 1
QSM 0.8 × 0.8 × 2 12° Min 40 Multi
7.0 T SWI 0.3 × 0.3 × 1.2 15° 15 28 1
QSM 0.8 × 0.8 × 0.8 Min 34 Multi

These sequence parameters are not intended as general “recommendations.” Rather, they serve to illustrate parameter examples that produce good SWI and QSM results. The parameters should be modified for different purposes; for example, multiple echoes can also be acquired for SWI.