Skip to main content
. 2015 Apr 24;42(5):2585–2593. doi: 10.1118/1.4918581

TABLE IV.

RMSE, maximum difference, and standard deviation between motion trajectories obtained from fluoroscopy of external radio-opaque markers and from corresponding ROIs monitored by VisionRT.

ROI #1 (around external marker #1) ROI #2 (around external marker #2) ROI #3 (around external marker #3)
Patient respiratory profile RMSE (mm) Maximum difference (mm) STD (mm) RMSE (mm) Maximum difference (mm) STD (mm) RMSE (mm) Maximum difference (mm) STD (mm)
1 0.16 0.80 0.76 0.23 0.83 0.70 0.20 0.67 0.44
2 0.17 0.62 0.62 0.21 0.56 0.56 0.20 0.66 0.39
3 0.16 0.97 0.57 0.23 0.89 0.53 0.16 0.55 0.34
4 0.20 0.83 0.70 0.22 0.84 0.63 0.19 0.64 0.39
5 0.15 0.72 0.38 0.16 0.87 0.32 0.18 0.55 0.28
6 0.13 0.47 0.45 0.16 0.53 0.30 0.17 0.51 0.31
7 0.09 0.45 0.32 0.14 0.38 0.24 0.18 0.47 0.25
8 0.20 1.27 0.61 0.23 0.78 0.55 0.19 0.73 0.36
9 0.26 1.48 1.30 0.24 0.97 1.22 0.21 0.68 0.66
10 0.30 1.33 1.01 0.26 0.90 0.94 0.23 0.69 0.57