Table 1.

HPLC method verification results for insulin assay.

Analytical Parameter Result
Linearity Range
- 0.9–10.0 mg/mL R2 = 0.9998
Specificity
- Interference No interference from formulation components and related compounds
Precision (%)
- Inter-day RSD (n = 3 × 3) 0.8–1.1
- Intra-day RSD (n = 5) 0.7–0.9
Recovery (%) 98.3 ± 0.9
DL (mg/mL) 0.02
QL (mg/mL) 0.08
Stability of Solution
- Decrease in assay at ambient condition 2.5% decrease in assay after 24 h
System Suitability
- Resolution between insulin and A-21 desamido insulin 3.7
- Tailing factor for insulin peak <1.8
- RSD of replicate injections <1.6%

DL and QL: detection and quantitation limits, RSD: relative standard deviation.