Table 3. Comparison of the GAT, RT, and DCT derived IOP readings.
Parameters (mm Hg) | All subjects n=124 | Normal n=56 | All glaucoma n=68 | |||
GAT | 16.00±3.80 | 16.14±3.81 | 15.88±3.82 | |||
Dif | P | Dif | P | Dif | P | |
RT | +0.99 | 0.002 | +1.75 | <0.001 | +0.37 | 0.563 |
DCT | +4.40 | <0.001 | +4.06 | <0.001 | +4.69 | <0.001 |
GAT: Goldmann applanation tonometer; RT: Rebound tonometer; DCT: Dynamic contour tonometer; IOP: Intraocular pressure; Dif: Difference from GAT readings.