Skip to main content
. 2015 Apr 18;8(2):299–304. doi: 10.3980/j.issn.2222-3959.2015.02.15

Table 6. Results of Bland-Altman analyses of the agreement between GAT, RT, and DCT.

Parameters x±s difference (mm Hg) 95% LoA (mm Hg)
All subjects
 GAT-RT -0.99±3.25 -7.4 to 5.4
 GAT-DCT -4.40±3.26 -10.8 to 2.8
 RT-DCT -3.41±3.10 -2.7 to 9.5
Normal eyes
 GAT-RT -1.75±3.41 -8.4 to 4.9
 GAT-DCT -4.06±3.42 -10.8 to 2.6
 RT-DCT -2.31±3.16 -3.9 to 8.5
Glaucomatous eyes
 GAT-RT -0.37±3.00 -6.3 to 5.5
 GAT-DCT -4.67±3.12 -10.8 to 1.4
 RT-DCT -4.32±2.77 -1.1 to 9.7

GAT: Goldmann applanation tonometer; RT: Rebound tonometer; DCT: Dynamic contour tonometer; LoA: Limits of agreement.