Table 2.

Comparison of total retinal thickness measurements in 9 macular sectors obtained using different optical coherence tomography instruments (mean (SD) of thickness measurements).

Spectralis, HRA, Heidelberg Engineering, Germany Spectralis, Heidelberg Engineering, Germany Cirrus HD-OCT, Carl Zeiss Meditec, Inc.** RTVue-100, Optovue, Inc., Fremont, CA** 3D OCT-1000, Topcon, Inc., Paramus, NJ** 3D OCT-1000, Topcon, Inc.†† Stratus OCT [OCT3], Carl Zeiss Meditec, Dublin, CA‡‡

Macular sector
Fovea 265 (15)* 264 (13) 262 (16) 256 (15) 231 (16) 222 (18) 212 (20)
Parafovea
 Temporal 307 (15) 306 (14) 306 (10) 308 (13) 280 (10) 285 (14) 251 (13)
 Superior 312 (13) 311 (14) 320 (12) 324 (11) 293 (12) 297 (15) 255 (17)
 Nasal 325 (16) 325 (15) 323 (12) 324 (11) 296 (12) 299 (15) 267 (16)
 Inferior 316 (14) 314 (13) 316 (11) s318 (10) 288 (10) 294 (15) 260 (15)
Perifovea
 Temporal 255 (11) 256 (14) 255 (9) 265 (10) 234 (16) 244 (12) 210 (14)
 Superior 273 (17) 271 (18) 274 (13) 278 (13) 249 (13) 257 (13) 239 (16)
 Nasal 272 (17) 298 (19) 293 (13) 291 (14) 266 (13) 273 (14) 246 (14)
 Inferior 265 (16) 274 (17) 264 (11) 267 (12) 240 (12) 247 (13) 210 (13)

*Mean (SD) of thickness measurements are shown in μm.

Current study.

Loduca and associates [34].

**Sull and associates [6].

††Ooto and associates [41].

‡‡Chan and associates [10].