Skip to main content
. Author manuscript; available in PMC: 2016 Apr 1.
Published in final edited form as: Anal Bioanal Chem. 2015 Feb 19;407(10):2777–2789. doi: 10.1007/s00216-015-8508-6

Table 1.

Estimation of the MALDI-TOF MS minimal detectable concentration for CTB bound to 1 mol% GM1 in poly(bis-SorbPC) PSLBs.a

Concentration of CTB (nM) Frequency of samples that gave a S/N ≥ 3b Number of silicon wafers prepared
1 0/10 2
2 8/16 3
3 11/15 3
4 9/9 2
5 10/10 2
a

This table focuses on a narrow CTB concentration range that was identified from screening samples over a larger concentration range.

b

Frequency is reported as the number of samples in which CTB was detected/total samples that were analyzed.