Table 3.

Comparisons of measured head scatter factor with that of TG-74 published data at 10 cm water equivalent depth in mini phantoms.

Field size (cm2) 6MV-FB Siemens (Primus) TG-74 % of deviation 6MV-FB Varian (Clinac iX) TG-74 % of deviation
5 0.9614 0.9610 0.04 0.9642 0.9680 −0.40
10 1.0000 1.0000 0.00 1.0000 1.0000 0.00
15 1.0175 1.0170 0.05 1.0154 1.0160 −0.06
20 1.0253 1.0270 −0.17 1.0245 1.0260 −0.15
30 1.0317 1.0320 −0.03 1.0372 1.0410 −0.37
40 1.0326 1.0320 0.06 1.0476 1.0510 −0.32