Table 1.

QA tests for image guidance system.

Frequency QA test Tolerance
Daily Safety checks Functional
Imager positioning/repositioning ±1 mm
Imaging and treatment co-ordinate coincidence (single gantry angle) ±1 mm
Monthly Imaging and treatment co-ordinate coincidence (4 cardinal gantry angles) ±1 mm
Image scaling ±2 mm
Mechanical alignment – full range of travel ±2 mm
Geometric accuracy for CBCT ±2 mm
Automatic image registration and offset calculation accuracy ±2 mm
Image quality – planar and CBCT Baseline ± 5%
Annual Imaging dose Baseline ± 3%