Table 4.
Name | Sequence parameters |
---|---|
3D T1 MPRAGE | TR = 1.79 s, TE = 3.06 ms, TI = 1.050 s, FoV = 250 × 250 mm2, flip angle = 10°, voxel size = 1 × 1 × 1 mm3 |
pCASL 2D GE EPI | TR = 4 s, TE = 12 ms, FoV 220 × 220 mm2, flip angle = 90°, voxel size = 3.4 × 3.4 × 5.0 mm3, labeling duration 1.5 s, postlabeling delay 1.2 s, 40 label/control pairs |
T2 FLAIR | TR = 9 s, TE = 76 ms, TI = 2.5 s, FoV = 220 × 220 mm2, flip angle = 146°, voxel size = 0.9 × 0.9 × 2 mm3 |
T1 FLASH | TR = 300 ms, TE = 2.46 ms, FoV = 225 × 225 mm2, flip angle = 60°, voxel size = 0.9 × 0.9 × 3 mm3 |
GRE Field Mapping | TR = 499 ms, TE1 = 5.19 ms, TE2 = 7.65 ms, FoV = 192 × 192 mm2, flip angle = 60°, voxel size = 3 × 3 × 3 mm3 |
Resting fMRI/GE EPI | TR = 3 s, TE 30 ms, FoV = 192 × 192 mm2, flip angle = 90°, voxel size = 3 × 3 × 3 mm3 isotropic, 102 volumes |
DTI SE EPI | TR = 11 s, TE = 76.4 ms, b = 0 and 1000 s/mm2, 30 directions, FoV = 256 × 256 mm2, voxel size = 2 × 2 × 2 mm3 |
Proton Density/T2 MRI | TR = 6.22 s, TE1= 14 ms, TE2 = 98 ms, FoV = 256 × 256 mm2, flip angle = 150°, voxel size = 1 × 1 × 2 mm3 |
MPRAGE, magnetization prepared rapid acquisition gradient echo; pCASL, pseudocontinuous arterial spin labeling; GE EPI, gradient-echo echo-planar imaging; FLAIR, fluid-attenuated inversion recovery; FLASH, fast low angle shot; GRE, gradient recalled echo; fMRI, functional MRI; DTI, diffusion tensor imaging; PD, proton density; T1, T2, weighting of applied MRI sequence; TR, repetition time; TE, echo time; TI, inversion time; FoV, field of view.