Fig. 3.

Fig. 3

Electronics to process the micro-Hall signal. (a) The hall resistance R was measured versus the applied magnetic field B, and a sensitivity of 78 Ω/T was calculated. (b) The input-referred noise of the μHall sensors as measured with a spectrum analyzer. (c) A schematic of the electronics to condition the μHall signal prior to software analysis. Each pair of μHall sensors are driven with a current source. The differential output of each μHall sensor is AC coupled to a two-stage amplifier (gain = 30 × 30) before analog-to-digital conversion ADC. Once digitized the signal is sent to either a computer or mobile device for software analysis