Table 5. Alignments conducted between the MV X-ray beam isocenter and the CBCT image-guidance system, and a comprehensive evaluation of the manual image matching accuracy.

Source of uncertainty Gantry rotation Ring rotation CBCT setting CBCT registration U all
u 1 u 2 u 3 u 4
Divided factor 3 3 3 1
LR (mm) 0.20 0.20 0.55 0.22 0.84
SI (mm) 0.20 0.20 0.27 0.68 1.43
AP (mm) 0.20 0.20 0.08 0.25 0.60
Vector length (mm) 0.28 0.28 0.56 0.47 1.23

The alignments exhibited an overall variation of U all (95% confidence interval).