Figure 5.

Figure 5

An example segmentation error produced by the Cirrus software. (A) Erroneous outer segmentation line results in an erroneous mean retinal thickness measurement of 473 μm in superior surrounding area. (B) Accurate outer segmentation line after manual correction results in a mean retinal thickness of 873 μm. (C) Accurate outer segmentation line placed automatically by DOCTRAP software.