Skip to main content
. 2013 Nov 15;54(12):7595–7602. doi: 10.1167/iovs.13-11762

Table 2.

Mean Retinal Thickness Determined by DOCTRAP Automatic Software and Spectralis Semi-Automatic Software (Manual Placement of Inner RPE Boundary)


Retinal Thickness
Mean Thickness, μm
DOCTRAP Automatic Software vs. Spectralis Semi-Automatic Software
DOCTRAP Automatic Software, Mean ± SD*
Spectralis Semi-Automatic Software, Mean ± SD*
Mean Paired Difference, Mean ± SE
ICC (95% CI)
95% Limits of Agreement
1-mm diameter central foveal area 441.9 ± 107.9 443.9 ± 108.6 −2.02 ± 1.35 0.998 (0.996–0.9999) −4.84 to 0.80
Surrounding area, out to 3 mm
 Superior 424.3 ± 90.1 422.3 ± 88.1 1.99 ± 1.53 0.997 (0.993–0.999) −1.21 to 5.19
 Nasal 399.0 ± 68.8 396.4 ± 66.1 2.64 ± 1.79 0.993 (0.982–0.997) −1.12 to 6.39
 Inferior 389.3 ± 63.3 385.2 ± 58.5 4.14 ± 2.20 0.987 (0.967–0.995) −0.45 to 8.74
 Temporal 436.4 ± 103.9 435.5 ± 102.7 0.87 ± 1.50 0.998 (0.995–0.999) −2.26 to 4.01
*

n = 20.