Published in final edited form as: IEEE Trans Biomed Eng. 2015 Apr;62(4):1077–1088. doi: 10.1109/TBME.2014.2367233

TABLE III.

Geometric distortion evaluations under scan protocol T1W.

Line segment Actual distance (mm) Measured distance (difference %)
Baseline Robot running
ai 158.11 158.46(0.22) 158.39(0.17)
bj 150.00 150.46(0.31) 150.24(0.16)
ck 158.11 158.48(0.23) 158.03(0.05)
dl 141.42 141.51(0.07) 141.14(0.20)
em 158.11 157.97(0.09) 157.85(0.17)
fn 150.00 149.92(0.05) 149.89(0.07)
go 158.11 158.16(0.03) 158.24(0.08)
hp 141.42 141.65(0.16) 141.65(0.16)