Skip to main content
. 2015 May 1;15:69. doi: 10.1186/s12871-015-0039-3

Table 2.

Airway sealing pressure, cuff volume, intracuff pressure and pharyngolaryngeal morbidity

Variables Group P (P-LMA)(n = 40) Group G (G-LMA)(n = 40) Pvalue
Volume of air for cuff inflation to 60 cmH 2 O, ml 13.90 ± 1.19** 22.60 ± 3.27** <0.0001**
Airway sealing pressure, cmH 2 O 29.41 ± 4.33* 32.36 ± 4.93* 0.04*
Intracuff pressure measurement, cmH 2 O
30 min 58.83 ± 1.14 58.90 ± 1.15 0.823
60 min 58.67 ± 1.09 58.27 ± 1.20 0.182
90 min 58.40 ± 1.23 58.33 ± 1.35 0.870
120 min 58.27 ± 1.09 58.44 ± 1.03 0.658
Blood staining on removal of device 9/40 (22.5%) 8/40 (20%) 1.000
Postoperative pharyngolaryngeal morbidity
Sore throat
1 h 7/40 (17.5%) 6/40 (15%) 1.000
2 h 7/40 6/40
24 h 0 0
Dysphagia
1 h 5/40 (12.5%) 4/40 (10%) 1.000
2 h 5/40 4/40
24 h 0 0
Dysphonia
1 h 0 0
2 h 0 0
24 h 0 0

**P < 0.01, *p < 0.05, P-LMA: ProSeal laryngeal mask airway; G-LMA: Guardian laryngeal mask airway; h: hours, min: minutes, ml: milliliter.