Skip to main content
. 2015 May;40(3):264–267.

Table 1.

Prediction of ABG abnormalities by pulse oximetry in neonates with RDS

   Acidosis
(pH<7.20)
Hypercapnia
(PaCO2>60)
Hypoxemia
(PaO2<50)
Hyperoxemia
(PaO2>80)
Cut-off point SpO2 (%) 85 85 85 95
Sensitivity (%) 40 50 75 83
Specificity (%) 100 99.6 99.5 92.4
PPV (%) 100 75 75 83
NPV (%) 96.8 98.4 98.5 92.4

NPV: Negative predictive value; PPV: Positive predictive value