Skip to main content
. 2015 May 14;53(6):1938–1941. doi: 10.1128/JCM.00261-15

TABLE 3.

Performance of resolved test results for three NAATs in detecting GBS in antenatal specimens

Assay Sensitivity (95% CI) (%) Specificity (95% CI) (%)
AmpliVue 96.4 (86.6–99.3) 95.8 (90.8–98.3)
BD Max 100 (91.9–100) 95.8 (90.8–98.3)
illumigene 90.9 (79.3–96.6) 97.9 (93.6–99.5)