Skip to main content
. 2014 Jul 16;2(3):E135–E140. doi: 10.1055/s-0034-1377177

Table 1. Comparison of eCLE (endoscope-based CLE), pCLE (probe-based CLE) and SECM.

eCLE pCLE SECM
Image acquisition speed 0.8 frames/sec1 12 frames/sec2 100 frames/sec
Contrast mechanism Fluorescence Fluorescence Reflectance
Contrast agent Intravenous fluorescein Intravenous fluorescein Topical acetic acid
Implementation Integrated in endoscope Though an accessary port Over a guide wire
Laser wavelength 488 nm1 488 nm2 1300 nm
Resolution Lateral: 0.7 µm1 Axial: 7 µm1 Lateral: 1 µm2 Axial: 3 µm2 Lateral: 2 µmAxial: 17 µm
Imaging depth 0 – 250 µm1 0 – 50 µm3 86 – 114 µm4
Field of view (FOV) 500 × 500 µm2 1 240 × 200 µm2 2 280 × 1300 µm2
Contiguous mosaic size No published value available 8 mm2 5 1,100 mm2 6
Large-area imaging method Manual scanning Manual scanning Automatic pullback
1

These values were taken from Kiesslich et al. [1].

2

These values were taken from Becker et al. [6].

3

Imaging depth of pCLE can be set as a fixed value between 0 and 50 µm [16].

4

Multiple depth levels are imaged simultaneously over this range.

5

This value was taken from De Palma et al. [17].

6

This value is for a 5-cm-long pullback of the probe.