Skip to main content
. 2015 Jun;113:1–12. doi: 10.1016/j.neuroimage.2015.03.013

Table 1.

Comparison of tSNR values across all twelve experimental conditions. The mean tSNR values are reported with the standard deviation being over volunteers. Significant differences between the PMC on and PMC off cases are indicated by * for p < 0.05 and ** for p < 0.01. Mean and standard deviation of the partition-weighted integrated motion values are shown in parenthesis. The partition-weighted integrated motion values were not significantly different for any of the PMC On vs PMC off cases.

tSNR (partition-weighed integrated motion) No motion Slow motion Fast motion
3.0 mm case
PMC on 68.0 ± 2.9
(1.9 ± 0.3)
37.6 ± 6.6*
(8.4 ± 1.9)
23.3 ± 6.5*
(16.0 ± 9.0)
PMC off 63.6 ± 8.4
(2.3 ± 0.4)
27.3 ± 4.4
(6.5 ± 1.8)
17.4 ± 4.5
(15.3 ± 7.2)



1.5 mm case
PMC on 35.5 ± 4.3
(2.5 ± 0.7)
23.7 ± 2.5**
(8.1 ± 1.0)
18.0 ± 4.6*
(14.3 ± 9.0)
PMC off 33.3 ± 3.9
(1.8 ± 0.2)
16.6 ± 2.9
(8.2 ± 1.8)
13.6 ± 3.0
(14.5 ± 6.7)