Skip to main content
. 2015 May 26;10(5):e0127765. doi: 10.1371/journal.pone.0127765

Table 2. Performance of the automated stretching system on human bronchi.

Overall duration of cyclic stretching n Differences between measured and preset values
g %
Tbaseline
5 min 9 0.47±0.09** 178±28
First series 20 min 10 0.58±0.09** 164±12
40 min 10 0.52±0.09** 161±12
60 min 10 0.37±0.09** 156±21
Second series (validation) 60 min 79 0.42±0.05** 160±10
Tmax
5 min 9 -0.53±0.27* 87±10
First series 20 min 10 -0.19±0.29 96±11
40 min 10 0.00±0.21 109±8
60 min 10 0.23±0.26 115±10
Second series (validation) 60 min 79 0.06 ±0.06 106±3
Tension variation (Tmax—Tbaseline)
5 min 9 0.84±0.12** 43±9
First series 20 min 10 0.35±0.15* 52±9
40 min 10 0.15±0.22 58±12
60 min 10 0.13±0.21 88±21
Second series (validation) 60 min 79 0.14±0.13 55±5

Data are shown as the mean ± SEM. The imposed tension was measured every 0.1 sec with a dedicated software package (IOX, version 2.4.2, EMKA Technologies, Paris, France). The magnitude of the stretch generated by the system corresponded to the difference between 80% Tmax and Tbaseline (Fig 2). The expected tension variation generated by the device (80% Tmax—Tbaseline) was compared to the observed values generated by the bronchial rings (maximum tension—minimum tension) in response to the forces imposed by the device (Fig 2).

*P <.01

**P <.001 when comparing measured and preset values.