Table 2. Performance of the automated stretching system on human bronchi.
Overall duration of cyclic stretching | n | Differences between measured and preset values | ||
---|---|---|---|---|
g | % | |||
Tbaseline | ||||
5 min | 9 | 0.47±0.09** | 178±28 | |
First series | 20 min | 10 | 0.58±0.09** | 164±12 |
40 min | 10 | 0.52±0.09** | 161±12 | |
60 min | 10 | 0.37±0.09** | 156±21 | |
Second series (validation) | 60 min | 79 | 0.42±0.05** | 160±10 |
Tmax | ||||
5 min | 9 | -0.53±0.27* | 87±10 | |
First series | 20 min | 10 | -0.19±0.29 | 96±11 |
40 min | 10 | 0.00±0.21 | 109±8 | |
60 min | 10 | 0.23±0.26 | 115±10 | |
Second series (validation) | 60 min | 79 | 0.06 ±0.06 | 106±3 |
Tension variation (Tmax—Tbaseline) | ||||
5 min | 9 | 0.84±0.12** | 43±9 | |
First series | 20 min | 10 | 0.35±0.15* | 52±9 |
40 min | 10 | 0.15±0.22 | 58±12 | |
60 min | 10 | 0.13±0.21 | 88±21 | |
Second series (validation) | 60 min | 79 | 0.14±0.13 | 55±5 |
Data are shown as the mean ± SEM. The imposed tension was measured every 0.1 sec with a dedicated software package (IOX, version 2.4.2, EMKA Technologies, Paris, France). The magnitude of the stretch generated by the system corresponded to the difference between 80% Tmax and Tbaseline (Fig 2). The expected tension variation generated by the device (80% Tmax—Tbaseline) was compared to the observed values generated by the bronchial rings (maximum tension—minimum tension) in response to the forces imposed by the device (Fig 2).
*P <.01
**P <.001 when comparing measured and preset values.