Skip to main content
. 2015 May 26;10(5):e0128370. doi: 10.1371/journal.pone.0128370

Table 2. HerpeSelect Express POC test compared to HerpeSelect-2-ELISA.

Correlation of positive and negative findings, n = 89.

ELISA
negative n (%) positive n (%)
POC negative 76 (100.0) 10 (76.9)
positive 0 (0.0) 3 (23.1)
Sensitivity (95% CI): 23.1% (15.4%- 32.9%)
Specificity (95% CI): 100% (95.3%- 100%)
Concordance rate: 88.8%