Existing and proposed dwell time patterns in ring applicator with 4 cm intra-uterine (IU) tandem. IUTT/RT is 1 : 1 (existing approach) and 2 : 1 (proposed approach)

Dwell points From tip [cm] X-pos [cm] Y-pos [cm] Z-pos [cm] Existing attern Dwell time [s] Proposed pattern Dwell time [s]
Intra-uterine (4 cm) applicator
1 0.68 –0.02 –0.06 –3.44 69.17 85.49
2 1.18 –0.02 –0.06 –2.94 69.17 85.49
3 1.68 –0.02 –0.06 –2.44 69.17 85.49
4 2.18 –0.02 –0.06 –1.94 69.17 85.49
5 2.68 –0.02 –0.06 –1.44 69.17 85.49
6 3.18 –0.02 –0.06 –0.94 69.17 85.49
7 3.68 –0.02 –0.06 –0.44 69.17 85.49
8 4.18 –0.02 –0.06 0.06 69.17 85.49
553.38 (9.22 min) 683.95 (11.40 min.)
Right ring (30 mm) applicator
1 0.68 1.43 0.26 0.21 138.34 85.49
2 1.18 1.45 0.76 0.23 138.34 85.49
276.69 (4.62 min) 170.99 (2.85 min)
Left ring (30 mm) applicator
1 0.68 –1.55 0.34 0.03 138.34 85.49
2 1.18 –1.52 0.84 0.05 138.34 85.49
276.69 (4.62 min) 170.99 (2.85 min)