Published in final edited form as: Med Phys. 2003 Jul;30(7):1671–1682. doi: 10.1118/1.1578911

Table I.

Tracking experiments. The first two columns represent the mean distance that each corresponding ball moved between trials. The last column presents the difference between the range scanner and OPTOTRAK distance measurements of corresponding fiducials. The depth range spanned most of the scanner’s working field of view, with a minimum depth of 493.1 mm and a maximum of 1035.6 mm.

Mean distance of fiducial movement between sets
Difference of corresponding fiducial distances between range scanner and OPTOTRAK (max)
Range scan, mm OPTOTRAK, mm
Near to middle 154.0±0.6 154.0±0.4 0.4±0.2 (0.8)
Middle to far 220.7±5.7 220.7±5.6 0.4±0.3 (1.0)
Near to far 374.6±5.4 374.6±5.2 0.8±0.6 (1.7)