Skip to main content
. Author manuscript; available in PMC: 2016 Sep 1.
Published in final edited form as: Clin Neurophysiol. 2014 Nov 28;126(9):1790–1796. doi: 10.1016/j.clinph.2014.11.017

Table 4b.

50/200 kHz phase ratio data separated for DMD and healthy boys.

B. 50/200 Ratio Healthy Healthy Healthy DMD DMD DMD

Intra-evaluator N % RSD ICC N % RSD ICC
Long. 6-Muscle 19 2.2 ± 1.8 0.92 12 3.0 ± 1.8 0.91
Long. 2-Muscle 21 2.9 ± 1.9 0.90 12 3.6 ± 3.6 0.81
Trans. 6-Muscle 17 2.4 ± 1.7 0.93 11 4.2 ± 3.3 0.81
Trans. 2-Muscle 20 3.0 ± 1.7 0.93 12 4.5 ± 3.4 0.78

Inter-evaluator
Long. 2-Muscle 20 2.9 ± 2.1 0.89 12 3.5 ± 3.0 0.85
Trans. 2-Muscle 19 2.7 ± 2.2 0.93 12 3.6 ± 3.1 0.87