Published in final edited form as: Clin Chim Acta. 2015 Apr 8;446:54–60. doi: 10.1016/j.cca.2015.03.025

Table 1.

Imprecision of Capillarys 2 Flex Piercing HbA1c assay.

NGSP units
% HbA1c
Mean
Imprecision (CV, %)
Center 1 Center 2 Inter-
laboratory
Within-
run
Between-run Total Within-
run
Between-
run
Total Total
4.7 2.2 0.0 2.3 2.0 0.0 2.0 2.1
6.3 1.2 0.6 1.4 1.4 0.4 1.5 1.4
7.4 1.2 0.7 1.4 1.3 0.0 1.3 1.3
11.1 1.1 0.5 1.2 0.9 0.4 1.0 1.1
SI units
HbA1c
Mean
(mmol/mol)
Imprecision (CV, %)
Center 1 Center 2 Inter-
laboratory
Within-
run
Between-run Total Within-
run
Between-
run
Total Total
28 3.7 0.0 3.8 3.4 0.9 3.5 3.7
45 1.9 0.7 2.0 2.1 0.7 2.2 2.1
57 1.7 0.9 2.0 1.9 0.0 1.9 1.9
98 1.3 0.5 1.4 1.0 0.5 1.2 1.3