Skip to main content
. 2015 May 27;10:119. doi: 10.1186/s13014-015-0424-6

Table 1.

Ultrasound probe orientation affected mostly the longitudinal position readings

Number of experiment, number of observer Slope in longitudinal direction (mm/°) Slope in lateral direction (mm/°) Slope in vertical direction (mm/°)
1st experiment, 1st observer (HB) +0.123 ± 0.006 (r2 = 0.92) –0.016 ± 0.005 (r2 = 0.19) +0.004 ± 0.002 (r2 = 0.06)
1st experiment, 2nd observer (BDB) +0.122 ± 0.009 (r2 = 0.82) –0.015 ± 0.004 (r2 = 0.30) – 0.010 ± 0.004 (r2 = 0.16)
1st experiment, 3rd observer (MR) +0.126 ± 0.005 (r2 = 0.95) –0.013 ± 0.005 (r2 = 0.16) +0.001 ± 0.003 (r2 = 0.00)
2nd experiment, 1st observer (HB) +0.211 ± 0.009 (r2 = 0.93) –0.017 ± 0.008 (r2 = 0.12) +0.026 ± 0.009 (r2 = 0.18)
2nd experiment, 2nd observer (BDB) +0.157 ± 0.008 (r2 = 0.91) +0.014 ± 0.006 (r2 = 0.12) +0.035 ± 0.006 (r2 = 0.43)
3rd experiment, 1st observer (HB) +0.124 ± 0.006 (r2 = 0.93) +0.013 ± 0.004 (r2 = 0.21) -0.016 ± 0.003 (r2 = 0.48)
3rd experiment, 2nd observer (BDB) +0.110 ± 0.005 (r2 = 0.92) +0.013 ± 0.003 (r2 = 0.42) -0.011 ± 0.003 (r2 = 0.29)