Skip to main content
. Author manuscript; available in PMC: 2015 Jun 1.
Published in final edited form as: IEEE Trans Med Imaging. 2005 Jul;24(7):868–877. doi: 10.1109/tmi.2005.852501

TABLE V.

Measured SNRs and Contrasts, at 2nd OSEM Iteration, Smoothed With a Hann Filter

Orbit Lesion Size
9.5 mm (0.43 ml) 8 mm (0.26 ml) 5 mm (0.06 ml)
SNR Contrast SNR Contrast SNR Contrast
Ideal low-noise case with high uptake:
935 ml Breast Only
10:1 activity concentration ratio, 12.1 μCi/ml in lesions, 16% EWin

central distal

TPB 19.3 2.0 2.7 0.2
CP2A 26.6 1.9 4.9 0.2
CPA 17.3 1.4 3.9 0.2
CLOVER 23.2 1.4 7.9 0.3
invCLOVER 17.1 1.5 6.9 0.3

More realistic case simulating patient biology:
1060 ml Breast + Filled Torso,
6.1:1:12 activity concentration ratio, 2.0 μCi/ml in lesions, 8% EWin

central distal

TPB 4.6 0.8 3.4 0.5
CPA 4.8 1.0 5.5 0.6
CLOVER 3.7 0.8 2.0 0.3
invCLOVER 6.3 1.2 1.2 0.2