Table 1. The mean spherical equivalent (SE) and Jackson cross cylinder power values obtained by the Plusoptix A09 and cycloplegic retinoscopy.
Mean SE value (D) | Mean cylinder power value J0 (D) | Mean cylinder power value J45 (D) | |
---|---|---|---|
The Plusoptix A09 | 1.14 ± 2.19 | 0.05 ± 0.63 | -0.03 ± 0.70 |
Cycloplegic retinoscopy | 1.71 ± 2.52 | -0.03 ± 0.51 | 0.05 ± 0.58 |
P values * | 0.00 | 0.14 | 0.26 |
*Paired t test