Skip to main content
. 2015 Jun 1;10(6):e0126052. doi: 10.1371/journal.pone.0126052

Table 3. Sensitivity and specificity for detecting refractive amblyopia risk factors with the Plusoptix A09 according to revised criteria on the basis of receiver operating characteristic (ROC) curve.

Revised criteria Hyperopia>1.88 D in any meridian Myopia>-3.0 D in any meridian Astigmatism (>1.5 D within 10°of 90°or180°,>1.0 D in oblique axis) Anisometropia >1.25 D (spherical or cylindrical)
Sensitivity 86.7% 85.7% 85.0% 72.2%
Specificity 89.5% 94.7% 85.5% 84.8%