Table 3. Sensitivity and specificity for detecting refractive amblyopia risk factors with the Plusoptix A09 according to revised criteria on the basis of receiver operating characteristic (ROC) curve.
Revised criteria | Hyperopia>1.88 D in any meridian | Myopia>-3.0 D in any meridian | Astigmatism (>1.5 D within 10°of 90°or180°,>1.0 D in oblique axis) | Anisometropia >1.25 D (spherical or cylindrical) |
---|---|---|---|---|
Sensitivity | 86.7% | 85.7% | 85.0% | 72.2% |
Specificity | 89.5% | 94.7% | 85.5% | 84.8% |