Skip to main content
. 2015 May 19;2015:571473. doi: 10.1155/2015/571473

Table 2.

PET/CT scanners: reconstruction parameters.

Discovery
ST
S1
Discovery
ST
S2
Discovery STE
S3
Discovery
600
S4
Discovery
600
S5
Discovery
690
S6
Biograph
16 Hi-REZ
S7
Biograph 6
true V
S8
Gemini
XL
S9
Gemini
TF
S10
Gemini
TF
S11
Reconstruction protocol FORE-OSEM FORE-OSEM 3D-OSEM 3D-OSEM 3D-OSEM 3D-OSEM FORE-OSEM FORE-OSEM LOR-RAMLA TF-MLEM TF-MLEM
Number of iterations 14, 42 16, 40 14, 42 16, 32 16, 48 54, 108 16, 24 16, 42 66 60, 99 60, 99
Transaxial smoothing
FWHM (mm)
6, 9, 13 5.5, 8.2, 11 6, 8, 11 6, 8, 11 5.5, 8.2, 11 4, 6, 8 4, 6, 8 4, 6, 8
Axial smoothing
(kernel)
1 : 4 : 1 1 : 4 : 1 1 : 4 : 1 1 : 6 : 1  
1 : 4.1  
1 : 2 : 1
Kernel width (cm) 14.1 14.1, 18.7
Relaxation (λ) 0.5, 0.7, 1 0.5, 0.7, 1 0.5, 0.7, 1
Voxel dimensions
L × W × H (mm)
2.7 × 2.7 × 3.3 2.7 × 2.7 × 3.3 2.7 × 2.7 × 3.3 2.7 × 2.7 × 3.3 2.7 × 2.7 × 3.3 2.7 × 2.7 × 3.3 2.6 × 2.6 × 2  
5.3 × 5.3 × 2
4.1 × 4.1 × 5 4 × 4 × 3 4 × 4 × 3 4 × 4 × 3
Number of reconstructions 6 6 6 6 6 18 12 6 3 6 9
Number of data points 1641 1651 1676 1785 1727 5368 3456 1676 814 1993 2466

Only available with 99 iterations.