. 2014 Aug 18;87(1041):20140240. doi: 10.1259/bjr.20140240

Table 1.

Means and standard deviations (SDs) of the skull-positioning errors measured immediately after treatment

Direction Fraction number
All fractions 2.11 SD
1 2 3 4
x (mm) −0.05 ± 0.47 −0.12 ± 0.42 0.16 ± 0.44 −0.04 ± 0.49 −0.01 ± 0.46a 0.97
y (mm) −0.07 ± 0.75 −0.11 ± 0.72 −0.23 ± 0.60 −0.02 ± 0.45 −0.11 ±  0.63a 1.30
z (mm) −0.16 ± 0.54 −0.06 ± 0.38 −0.06 ± 0.41 −0.23 ± 0.29 −0.13 ± 0.42a 0.88
Rx (deg) 0.10 ± 0.42 0.18 ± 0.36 0.06 ± 0.29 0.15 ± 0.29 0.12 ± 0.34  
Ry (deg) 0.15 ± 0.38 −0.08 ± 0.51 −0.08 ± 0.41 −0.1 ± 0.35 −0.03 ± 0.42  
Rz (deg) −0.03 ± 0.44 −0.19 ± 0.68 −0.22 ± 0.60 −0.33 ± 0.56 −0.19 ± 0.57  

Each error was calculated by the bone-matching procedure available in the on-board cone beam CT system, and therefore, considered as an average positioning error during beam delivery. Mean errors in all fractions are also shown with SDs.

a

Planning target volume margins were calculated by 2.11 times the SD in x, y and z directions.