Classification accuracy and sensitivity values using EEG measures and hemodynamic parameters during cardiac surgery for all patients (n = 25)

Method SEN (%) (awake) SEN (%) (Light anesthetic) SEN (%) (surgical anesthetic) SEN (%) (deep anesthetic) Classification accuracy (%)
HR parameter 45.2 27.3 12.2 48.4 33.4 ± 7.6
MAP parameter 53.2 29.3 11.4 48.3 35.6 ± 7.8
HR and MAP parameters 61.2 40.4 26.3 85.4 53.5 ± 6.5
PE index 85.5 74.2 82.5 25.4 66.9 ± 5.4
MMPE index 87.5 81.2 88.5 83.4 85.1 ± 3.5
BIS index 93.5 65.6 87.8 75.6 80.6 ± 4.1
BIS index, HR and MAP 91.8 76.4 88.3 82.6 84.7 ± 3.6
MMPE index, HR and MAP 92.5 86.5 90.2 88.3 89.4 ± 3.2