Skip to main content
. 2015 Jun 3;5:10842. doi: 10.1038/srep10842

Figure 2. Probe limit of detection.

Figure 2

By measuring a range of magnetic tracer quantities at a fixed distance of 4.0 mm from the signal source, a 50 μg limit of detection (SNR = 2.0) was calculated. Due to the short-range nature of magnetic fields, this value can be significantly improved if the probe is moved closer to the signal source. Inline graphic = Measured data; Inline graphic = Calculated datum.