Sensitivity and specificity of MiniLab® tests for detecting drugs noncompliant with HPLC content analysis (frequency, proportion, and 95% exact confidence interval)

Antibiotic Test Sensitivity Specificity
Amoxicillin MiniLab® CT 0/1 19/19
0% [0.0–97.5] 100% [82.3–100.0]
MiniLab® TLC 0/1 17/19
0% [0.0–97.5] 89.5% [66.9–98.7]
Co-trimoxazole MiniLab® CT 0/14 1/1
0% [0.0–23.2] 100% [0.0–23.2]
MiniLab® TLC 2/14 1/1
14% [1.8–42.8] 100% [2.5–100.0]