Table 1.
Glucose meter results fulfilling error limits at: |
|||||||
---|---|---|---|---|---|---|---|
Device number | Glucose meter | n | Overall accuracy (%) | n | BG concentration < 100 mg/dl (%) | n | BG concentration ≥ 100 mg/dl (%) |
1 | Accu-Chek Aviva Nano | 303 | 97.0 | 35 | 91.4 | 268 | 97.8 |
2 | Accu-Chek Compact | 282 | 99.6 | 41 | 100.0 | 241 | 99.6 |
3 | Accu-Chek Mobile | 294 | 99.3 | 47 | 97.9 | 247 | 99.6 |
4 | BG Star | 302 | 90.1 | 44 | 90.9 | 258 | 89.9 |
5 | iBG Star | 269 | 97.4 | 44 | 97.7 | 225 | 97.3 |
6 | Breeze | 249 | 86.7 | 39 | 92.3 | 210 | 85.7 |
7 | Contour Plasma | 284 | 94.7 | 59 | 96.6 | 225 | 94.2 |
8 | Contour USB | 248 | 96.4 | 34 | 100.0 | 214 | 95.8 |
9 | Contour USB next | 275 | 99.6 | 37 | 97.3 | 238 | 100.0 |
10 | Contour XT | 288 | 99.7 | 41 | 100.0 | 247 | 99.6 |
11 | FineTouch | 292 | 93.5 | 41 | 80.5 | 251 | 95.6 |
12 | FreeStyle Lite | 301 | 97.7 | 42 | 100.0 | 259 | 97.3 |
13 | GL 40 | 254 | 94.9 | 37 | 97.3 | 217 | 94.5 |
14 | GL 50 | 251 | 86.5 | 32 | 93.8 | 219 | 85.4 |
15 | Glucomen LX | 251 | 92.0 | 34 | 82.4 | 217 | 93.5 |
16 | Glucomen LX plus | 246 | 91.5 | 43 | 93.0 | 203 | 91.1 |
17 | GlucoSmart Swing | 253 | 91.7 | 34 | 79.4 | 219 | 93.6 |
18 | my glucohealth | 285 | 91.9 | 45 | 95.6 | 240 | 91.3 |
19 | mylife Pura | 276 | 97.5 | 46 | 100.0 | 230 | 97.0 |
20 | Omnitest 3 | 252 | 80.2 | 36 | 86.1 | 216 | 79.2 |
21 | One Touch Ultra easy | 284 | 96.8 | 44 | 97.7 | 240 | 96.7 |
22 | One Touch Verio Pro | 292 | 92.8 | 31 | 83.9 | 261 | 93.9 |
23 | One Touch Verio IQ | 283 | 96.8 | 40 | 97.5 | 243 | 96.7 |
24 | One Touch Vita | 296 | 96.3 | 49 | 98.0 | 247 | 96.0 |
25 | Smart Lab Mini | 270 | 87.4 | 42 | 81.0 | 228 | 88.6 |
26 | Smart Lab Sprint | 280 | 87.1 | 33 | 84.8 | 247 | 87.4 |
27 | Wellion Calla | 251 | 91.6 | 36 | 77.8 | 215 | 94.0 |
Results fulfilling accuracy requirements are given in bold.