Skip to main content
. 2014 May;8(3):466–472. doi: 10.1177/1932296814522804

Table 2.

Accuracy Results of GMS After Differentiation in 5 BG ranges.

BG 50-99 mg/dl
BG 100-149 mg/dl
BG 150-199 mg/dl
BG 200-249 mg/dl
BG 250-300 mg/dl
Device number Glucose meter Data pairs n Max 15 mg/dl (%) Data pairs n Max 15% (%) Data pairs n Max 15% (%) Data pairs n Max 15% (%) Data pairs n Max 15% (%)
1 Accu-Chek Aviva Nano 35 91.4 53 96.2 60 98.3 86 96.5 69 100.0
2 Accu-Chek Compact 41 100.0 36 97.2 68 100.0 68 100.0 69 100.0
3 Accu-Chek Mobile 47 97.9 55 100.0 52 100.0 81 98.8 59 100.0
4 BG Star 44 90.9 65 89.2 66 89.4 60 90.0 67 91.0
5 iBG Star 44 97.7 70 100.0 56 94.6 56 98.2 43 95.3
6 Breeze 39 92.3 36 80.6 58 82.8 69 84.1 47 95.7
7 Contour Plasma 59 96.6 48 91.7 47 93.6 70 94.3 60 96.7
8 Contour USB 34 100.0 46 97.8 49 91.8 57 98.2 62 95.2
9 Contour USB next 37 97.3 70 100.0 71 100.0 53 100.0 44 100.0
10 Contour XT 41 100.0 66 100.0 73 98.6 66 100.0 42 100.0
11 FineTouch 41 80.5 57 100.0 65 96.9 75 93.3 54 92.6
12 FreeStyle Lite 42 100.0 62 95.2 59 93.2 78 100.0 60 100.0
13 GL 40 37 97.3 51 92.2 41 90.2 61 98.4 64 95.3
14 GL 50 32 93.8 64 85.9 56 91.1 49 87.8 50 76.0
15 Glucomen LX 34 82.4 50 90.0 44 95.5 56 91.1 67 97.0
16 Glucomen LX plus 43 93.0 57 94.7 57 93.0 42 92.9 47 83.0
17 GlucoSmart Swing 34 79.4 51 84.3 52 94.2 74 97.3 42 97.6
18 my glucohealth 45 95.6 68 91.2 76 88.2 60 96.7 36 88.9
19 mylife Pura 46 100.0 51 96.1 53 98.1 84 97.6 42 95.2
20 Omnitest 3 36 86.1 34 82.4 62 77.4 70 74.3 50 86.0
21 One Touch Ultra easy 44 97.7 63 95.2 54 96.3 72 98.6 51 96.1
22 One Touch Verio IQ 40 97.5 68 92.6 70 97.1 68 98.5 37 100.0
23 One Touch Verio Pro 31 83.9 60 78.3 64 96.9 67 98.5 70 100.0
24 One Touch Vita 49 98.0 69 95.7 61 96.7 78 96.2 39 94.9
25 Smart Lab Mini 42 81.0 54 83.3 58 94.8 73 83.6 43 95.3
26 Smart Lab Sprint 33 84.8 52 78.8 64 82.8 62 88.7 69 97.1
27 Wellion Calla 36 77.8 43 90.7 54 98.1 72 94.4 46 91.3
Mean (± SD) of data pairs 40 ± 6 56 ± 11 59 ± 9 67 ± 11 53 ± 11
Percentage of all data pairs 14.5 20.4 21.4 24.4 19.3

Results fulfilling accuracy requirements are given in bold.