Skip to main content
. 2014 Nov;8(6):1081–1090. doi: 10.1177/1932296814538940

Table 3.

Criteria Published by Different Institutions for Approval of Point-of-Care Blood Glucose Meters.

Year Standard setting institution and their guideline document Error margins Specifically suggested for blood glucose meters in hospital setting
1987 ADA Acceptable error should be <10% for 100% of glucose results between 30-400 mg/dl No
1996 ADA Acceptable error should be <5% for 100% of glucose results between 30-400 mg/dl No (deemed unachievable by IFCC)
2003 ISO 15197 At ≤75 mg/dl, 95% of results should be within ± 15 mg/dl
At >75 mg/dl, 95% of results should be within ± 20%
No (currently adopted by the FDA in the US and CE in Europe for hospital blood glucose meter certification)
2013 ISO 15197 At <100 mg/dl, 95% of results should be within ± 15 mg/dl
At ≥100 mg/dl, 95% of results should be within ± 15%
99% of individual results within zones A and B of the consensus error grid (error grid not used in 2003 ISO criteria)
No
2013 CLSI POCT 12-A3 (its previous version had adopted the ISO 15197:2003 criteria) At <100 mg/dl, 95% of results should be within ± 12 mg/dl
At ≥100 mg/dl, 95% of results should be within ± 12.5%
At ≤75 mg/dl, 98% of results should be within ± 15 mg/dl
At >75 mg/dl, 98% of results should be within ± 20%
Blood glucose meters should be evaluated in the populations where they intend to be used
Yes