Table 3. Comparison of the Macular Thickness of 9 Sectors on ETDRS Measured by the Topcon Algorithm and the Graph-Based Algorithm (Mean ± SD).

Normal subjects Glaucoma subjects
Topcon Graph-Based Correlation coefficient Mean Difference (Topcon-Graph Based) p-value Topcon Graph-Based Correlation coefficient Mean Difference (Topcon-Graph Based) p-value
Cen 223.6±18.5 217.3±17.8 0.95 6.3±5.9 <0.01 225.1±21.6 216.8±20.0 0.84 8.3±11.8 <0.01
Sin 299.7±17.0 296.5±17.2 0.96 3.2±4.5 <0.01 284.7±22.4 281.7±22.7 0.93 3.0±8.7 <0.01
Nin 300.9±18.0 296.7±17.4 0.93 4.2±6.5 <0.01 285.2±23.2 283.0±23.8 0.90 2.2±11.0 >0.06
Iin 295.7±17.0 293.7±16.4 0.94 2.1±5.7 <0.01 276.5±25.1 275.2±24.3 0.95 1.3±7.8 >0.12
Tin 284.0±16.5 283.6±16.3 0.95 0.4±5.3 >0.17 269.1±21.3 267.0±21.4 0.91 2.1±8.9 >0.02
Sout 265.7±15.0 262.8±15.5 0.94 2.8±5.5 <0.01 246.5±20.3 245.3±20.1 0.93 1.2±7.8 >0.12
Nout 281.8±16.6 284.3±21.3 0.86 -2.5±11.4 <0.01 263.2±22.9 263.4±26.2 0.71 -0.2±19.0 >0.90
Iout 254.0±14.2 251.6±15.2 0.90 2.5±6.7 <0.01 235.5±21.2 232.1±20.8 0.92 3.4±5.6 <0.01
Tout 249.7±14.5 246.2±15.4 0.79 3.6±9.7 <0.01 234.3±19.1 234.9±22.1 0.70 -0.6±16.2 >0.70
Ave 268.9±13.5 267.0±14.1 0.96 1.9±4.0 <0.01 251.4±18.6 250.7±19.0 0.93 1.0±7.4 >0.20