Table 4.
The volume estimations of glass beads imaged by a spin echo sequence. The first column lists the diameters of glass beads measured by a vernier caliper. The second column lists the pseudo volume V1. The third column lists the pseudo volume V2. The fourth column shows the estimated volume V0 from spin echo images. The fifth column is the calculated volume based on the first column. The sixth column is the difference in percentage between V0 and V0′. Accurate volume measurements of beads without MR signal can be obtained from spin echo images.
| Diameters of | V1 | V2 | V0 | V0′ | Diff |
|---|---|---|---|---|---|
| glass beads (mm) | (mm3) | (mm3) | (mm3) | (mm3) | % |
| 2.06 ± 0.05 | 320 | 57 | 4.4 ± 0.3 | 4.6 ± 0.3 | 4.3 |
| 3.08 ± 0.02 | 440 | 63 | 15.2 ± 0.3 | 15.3 ± 0.3 | 0.6 |
| 5.03 ± 0.10 | 1536 | 199 | 67.3 ± 0.4 | 66.6 ± 4.0 | 1.0 |
| 6.02 ± 0.10 | 1678 | 378 | 109.0 ± 0.8 | 114.2 ± 5.7 | 4.5 |