Table 1.

Linearity of the LDH assay

tBHP (μM) Slope R2 y-intercept at x = 0 n (x), n (y)
0 0.6634 ± 0.0013 0.9996 0.1275 ± 0.0023 91, 8
30 0.6769 ± 0.0014 0.9996 0.1317 ± 0.0025 91, 8
60 0.7129 ± 0.0017 0.9995 0.1369 ± 0.0030 91, 8
90 0.7634 ± 0.0022 0.9993 0.1464 ± 0.0038 91, 8
120 0.8417 ± 0.0028 0.9990 0.1604 ± 0.0049 91, 8
200 0.9848 ± 0.0058 0.9969 0.2064 ± 0.0100 91, 8

The linearity of the LDH assay was determined by linear regression analysis of the A490 absorbance values obtained during the first three hours of the assay at measurement intervals of 2 min. The assay showed a highly linear response (R2 > 0.99) throughout all conditions.