Abdominal CTC “standard-dose” protocol utilized in the prospective trial*

Protocol: Supine CTC
Scanner GE HD 750
Scan Type Helical
Rotation Time (sec) 0.5
Beam Collimation (mm) 40
Detector Rows 64
Pitch 0.984
Speed (mm/rot) 39.36
Detector Configuration 64 × 0.625
Slice Thickness for NI (mm) 1.25
Scan FOV Large Body
kV 120
Smart mA Range 30–300
Noise Index 50
Reconstructions (FBP):
  DFOV (cm) 36–50
  Recon Type Standard
  Window W/L 400/50
  Helical Recon Option Plus
  Slice Thickness (mm) 2.5
  Interval (mm) 1.5
*

The specific protocol for the accompanying low-dose series was derived by adjusting the noise index (NI)/slice thickness pairing (and mA range) to allow for a targeted 70–90% dose reduction (by DLP) relative to the “standard-dose” series.